Product Type:
Ventilator
Description:
The CARAT I pro is used to ventilate patients who are not yet fully dependent on mechanical ventilation. The device is suitable for the supply of adults as well as children from a respiratory volume of 50 ml. CARAT I pro was specially developed for home use, but is also suitable for use in hospitals. A choice of pressure- or volume-controlled ventilation is guaranteed via the setting of various ventilation modes.
Ventilation can be invasive (e.g. via a tracheostoma) or non-invasive (via a breathing mask). CARAT I pro is operated with a single-tube system. The device can be connected to a low-pressure gas source (with up to 15 l/min O2) for ventilation with increased oxygen concentration. It is also possible to combine CARAT I pro with a humidifier.
Contraindications:
- severe cardiac arrhythmias
- severe hypotension
- severe epistaxis
- pneumothorax or pneumomediastinum
- pneumoencephalus
- head trauma
- status post cranial or brain surgery
- acute sinusitis, otitis media or tympanic membrane perforation
- risk of aspiration
Features / Components:
- for life-sustaining and intermittent as well as invasive and non-invasive ventilation of adults and children (from a respiratory volume of 50 ml) with ventilatory insufficiency
- intended for operation with a single-tube system
- intended for mains and battery operation
- battery operation possible via internal rechargeable battery and external rechargeable battery pack
- 2-stage filter system with fine and coarse filter
- can be used in combination with a humidifier
- Delivery: consisting of basic unit with integrated rechargeable battery, Power supply unit and power cord, transport bag, reusable single hose system, adapter for hose system, spare filter cassette and spare filter as well as operating instructions
- external rechargeable battery available as optional accessory
Technical Data:
Weight: approx. 4.6 kg
Dimensions (W x D x H): 304 x 253 x 160 mm
Voltage: 100-250 V AC, 50-60 Hz, 24 V DC
Power consumption: approx. 60 W
Operating noise: approx. 25 dB(A) at 10 hPa
Therapy modes: PCV, PSV, SIMV, VCV
IPAP pressure range: 4 to 50 hPa
PS pressure range: 2 to 50 hPa
PEEP pressure range: 0 to 20 hPa
Minimum stable limit pressure: 0 hPa
Maximum stable limit pressure: 60 hPa
Minimum working pressure: 0 hPa
Maximum working pressure: 50 hPa
Breathing rate: 4 to 50 breaths per minute
Breathing pressure display: digital
Inspiratory time: 0.3 to 8 s
I:E ratio: 1:0.3 to 1:4.0
Trigger sensitivity: adjustable
Max. Flow: 180 l/min
Battery operation internal: 28.8 V, 2.3 Ah, lithium-ion battery
Battery operation external: 24 V, 7 Ah, AKKUPACK CARAT
Battery life: approx. 6 h (internal battery)
to 12 h (external battery)
Price (without guarantee):
The price is available on request from the manufacturer / distributor.