Kuvaus
Product Type:
Check valve to increase the assurance of oxygen equipment
Description:
FireSafe valves are specially designed for retrofitting oxygen concentrators. According to the amendment to DIN ISO standard 8359, which came into strength on July 01, 2012, in all oxygen concentrators the penetration of fire into the device at the oxygen outlet must be stopped and the oxygen supply must be interrupted in the event of ignition of the accessories close to the patient. For all older devices, retrofit options are to be made available on the part of the manufacturers. Retrofitting can also be performed as part of the annual maintenance.
Features / Components:
- thermal fuse designed to interrupt the oxygen supply if the upstream PVC hose line is ignited, thus stopping the fire
- plug-in connection between patient hose oxygen outlet on the device
- is attached directly to the oxygen outlet
Price (without guarantee):
The price is available on request from the manufacturer / distributor.