Product Type:
Ventilator
Indications:
- disorders of the respiratory center (e.g., multiple sclerosis or brainstem infarction)
- neuromuscular illnesses, e.g., spinal muscular atrophy or amyotrophic lateral sclerosis)
- disorders of transfer of muscle force to alveolar pressure (e.g., scoliosis)
- disorders of transfer of alveolar pressure to ventilation (e.g., airway obstruction)
- exhaustion of the respiratory pump (e.g. Cystic fibrosis)
Contraindications:
- after procedures on the thorax, heart, abdomen after facial surgery, consult with the surgeon
Features / Components:
- life-sustaining, invasive and non-invasive ventilation of adults and children (at least 5 kg) with ventilatory insufficiency
- ventilation with one and two-tube system
- ramp function for slow pressure increase during the sleep phase
- alarm function
- event history and trends are retrievable
- leakage test
- mains and battery operation
- battery operation possible via internal rechargeable battery and external battery pack
- complete monitoring and curves for pressure, Flow, volume and loop
- patient mode with security password
- TFT color display
- operation via rotary coding knob
- serial RS-232 interface
- Ethernet interface
- can be used in combination with the electric humidifier VAPOUR 2
- humidifier available as an optional accessory
Delivery:
- Basic unit
- Power supply unit
- Vehicle cable
- Reusable single hose system (1.8 m)
- Integrated rechargeable battery
- Filter set
- Operating instructions
Technical Data:
Weight: 3.9 kg
Dimensions (W x H x D): 24 x 21 x 33 cm
Width: 24 cm
Depth: 33 cm
Voltage: 230 V AC, 50 Hz; 12 V DC
Operating noise: approx. 35 dB
Therapy modes: Adaptive BiLevel, SIMV/PCV, SIMV/VCV, SIMV/PRVC,
A/C PCV, A/C VCV, A/C PRVC, CPAP/PSV
CPAP pressure range: 4 to 20 hPa
Pressure range Bilevel S,ST, PCV, APCV, PSV, PS.Tv 6 to 40 hPa
PEEP pressure range: 0 to 20 hPa
Tidal volume: 40 to 2500 ml
Respiratory rate: 5 to 50 RPM
Tidal volume: 50 to 2500 ml
Inspiratory time: 0.4 to 3 seconds
- min. Inspiratory time 0.3 to 2.5 seconds
- max. inspiratory time 1 to 3 seconds
inspiratory trigger: flow trigger, 1 to 9 l/min
expiratory trigger: flow trigger, 20 to 50% (relative to max. inspiratory flow)
max. flow: 120 l/min
Breathing pressure display: digital
Ramp duration: 0 to 50 min.
Ramp start pressure: 4 hPa to CPAP
Inspiratory time: 0.3 to 3 s
Apnea time: 5 to 60 sec
Max supplemental oxygen flow: 6 l/min
Battery life: up to 2.5 hr. (internal battery)
to 10 hrs. (external battery)
Price (without guarantee):
Price is available upon request from the manufacturer / distributor.