Product Type:
Ventilator
Description:
Pressure and volume controlled ventilator for continuous or intermittent respiratory support for patients requiring invasive or non-invasive mechanical ventilation.
Contraindications:
- Use of the product is contraindicated if a patient requires ventilation with a higher oxygen concentration (FiO2)
- Nasal mask ventilation therapy should be discussed with the surgeon after facial surgery.
Features / Components:
- ventilator for use in the clinic or at home
- for adult and pediatric patients (from 5kg)
- highly accurate ventilation volume and high trigger sensitivity
- extensive monitoring functions (SpO2, CO2, and FiO2)
- display of pressure, flow and volume
- measurement of exhaled volume with dual tube systems
- intuitive user interface
- selection of different tube systems possible-
Delivery:
- basic unit VIVO 60 incl. Power cord
- user manual
- carrying bag
- disposable filter, white, and washable filter, gray
- patient circuit with reusable leak valve and patient circuit with reusable active exhalation valve
- integrated rechargeable battery
Technical Data:
Dimensions (W x H x D): 348 × 120 × 264 mm without click-on battery
348 × 120 × 290 mm with click-on battery
Weight: 5.3 kg without click-on battery
6.7 kg with click-on battery
Mains voltage: 100 to 240 V AC
External battery: 24 V DC
Click-on battery: 8 hr.
Internal battery: 4 hrs.
Operating noise: below 30 dB(A)
Ventilation modes: PSV, PSV(TgV), MPV VCV, MPV PCV, SIMV PCV, PCV, PCV(TgV), SIMV VCV, PCV(A), PCV(A+TgV), CPAP, VCV, VCV(A), Sigh
Rise: 1 to 9 (PSV andPCV), 50 - 90%, Off (VCV)
Inspiratory pressure: 4 to 60 mbar
Inspiratory trigger: 1 to 9 (PSV, PCV andVCV, Off (PCV andVCV)
Expiratory trigger: 1 to 9 (PSV)
min. Inspiratory time: off, 0.3 to 3 s
max. Inspiratory time: 0.3 to 3 s, Off
Backup frequency (PSV): 4 to 60 bpm
Target volume: 50 to 2500 ml
Breath volume: 50 to 2500 ml
Flow pattern: rectangular, decelerating
Price (without guarantee):
Price is available upon request from the manufacturer/distributor.