Product Type:
Sleep apnea device
Indications:
- for obstructive sleep apnea and
high pressure requirement, poor compliance due to high expiratory pressure, fluctuating pressure requirement (position/REM sleep-dependent, night-to-night variation) or
isolated central parts
- for obesity hypoventilation syndrome (OHS) and insufficient therapeutic efficiency when using CPAP (CPAP failure) or poor compliance due to high expiratory pressure
Contraindications:
- cardiac decompensation
- Severe cardiac arrhythmia
- Severe hypotension, especially in conjunction with intravascular volume depletion
- Severe epistaxis
- High risk of barotrauma
- Serious lung disease (e.g. chronic obstructive pulmonary disease) E.g. chronic obstructive bronchitis (COPD), emphysema)
- hypoxemia during the day
- respiratory insufficiency of a cause other than OSA
- nocturnal hypoxemia of a cause other than OSA
- pneumothorax or pneumomediastinum
- pneumoencephalus
- head trauma
- status post brain surgery and post surgery on the pituitary gland or on the middle or inner ear Inner ear
- Acute sinusitis, otitis media or perforation of the eardrum
- Dehydration
Description:
Using patented OPP technology, the Auto-CPAP device can distinguish between central and obstructive respiratory disorders. It is also available with the softPAP exhalation aid. The powerful SOMNOaqua respiratory gas humidifier can be integrated as an option.
Features / components:
- for non-life-sustaining treatment of adults and children aged 12 and over
- Auto-CPAP therapy
- Click adapter for standard tubing
- 2-stage filter system with fine and coarse filter
- Can be used worldwide thanks to power supply unit with 100 to 240 V
- Mobile thanks to optional DC adapter for 12 or 24 V
- Soft start and freely adjustable, increased start pressure
- CPAP and APAP mode with optional softPAP exhalation facilitation
- OPP technology for differentiating between central and obstructive apneas
- remote adjustment with SOMNOadjust or WEINMANNsupport
- analog signals for PSG feed-in (pressure, flow, leakage, OPP, rel. AMV)
- can be used in combination with the SOMNOaqua heated humidifier
- humidifier either available as an optional accessory or included with the basic device
- suitable for mouth-nose masks
- trolley-compatible bag included
- NEW: Data transfer with SD memory card
Technical Data:
Weight: 1.7 kg
Dimensions (W x H x D): 21 x 9 x 27 cm
Connection diameter: 19.5 mm (breathing tube)
Voltage: 115/230 V AC, 50-60 Hz; 12/24 V DC
Power consumption at: 230 V 115 V 24 V 12 V
- Company 0.1 A 0.2 A 0.9 A 1.8 A
- Standby 0.02 A 0.04 A 0.2 A 0.4 A
Operating noise: approx. 27 dB(A) at 10 hPa
Operating pressure range: 4 to 18 hPa
Pressure accuracy: +/- 0.6 hPa
Max. CPAP pressure: <40 hPa
Flow at max. Speed and:
- 18 hPa 120 l/min
- 14 hPa 130 l/min
- 12 hPa 140 l/min
- 9 hPa 150 l/min
- 6 hPa 155 l/min
- 5 hPa 160 l/min
- 4 hPa 165 l/min
- 0 hPa 175 l/min
Tolerance: +/- 15 l/min
Fine filter separation efficiency: >= 99.5 % to 1 µm
>= 85 % to 0.3 µm
Fine filter service life: approx. 250 h
Order number: WM 27400 (white)
WM 27410 (anthracite)
HMV number: 14.24.21.0003
Price (without guarantee):